4.3 Displays in the standby mode

732 IC Detector / 733 IC Separation Center

81

4.3

Displays in the standby mode

4.3.1

Measured value and current time

In the standby mode of the conductivity measurement, the top line of

the display always shows the current measured value (left) and the

current time (right).

151.2

µµS/cm

13.5 min

method DEFAULT

Measured value
The following possibilities exist for the display of

the measured value:

Measured value without sign

151.2

µ

S/cm

Absolute conductivity in

µ

S/cm

or mS/cm.

This value is also shown auto-

matically if the auto-zero value

lies outside the full-scale range.

Measured value with sign

+0.004

µ

S/cm

Auto-zero value in

µ

S/cm

(default setting)

+0.0 %fs

Auto-zero value in %fs

(% of the full-scale range, can be

set under

"

>CONFIG/detector/

'zero' unit

"

)

+0.4 mV

Auto-zero value in mV

(can be set under

"

>CONFIG/

detector/'zero' unit

"

)

No measured value

--------

Measured value is more than

10% outside the set measuring

range.

151.2

µµS/cm

13.5 min

method DEFAULT

Current time
The following possibilities exist for the display of

the current time:

Without "INJECT A" or program start

Current time after instrument switched on
With "INJECT A"

Current time after the last switching of injection

valve A to "INJECT" position (does not apply to

an "INJECT A" command within a program)
With program start

Current time from last program start