3 displays in the standby mode, Measured value and current time, Status messages – Metrohm 732 IC Detector User Manual
Page 88
![background image](/manuals/586520/88/background.png)
4.3 Displays in the standby mode
732 IC Detector / 733 IC Separation Center
81
4.3
Displays in the standby mode
4.3.1
Measured value and current time
In the standby mode of the conductivity measurement, the top line of
the display always shows the current measured value (left) and the
current time (right).
151.2
µµS/cm
13.5 min
method DEFAULT
Measured value
The following possibilities exist for the display of
the measured value:
Measured value without sign
151.2
µ
S/cm
Absolute conductivity in
µ
S/cm
or mS/cm.
This value is also shown auto-
matically if the auto-zero value
lies outside the full-scale range.
Measured value with sign
+0.004
µ
S/cm
Auto-zero value in
µ
S/cm
(default setting)
+0.0 %fs
Auto-zero value in %fs
(% of the full-scale range, can be
set under
"
>CONFIG/detector/
'zero' unit
"
)
+0.4 mV
Auto-zero value in mV
(can be set under
"
>CONFIG/
detector/'zero' unit
"
)
No measured value
--------
Measured value is more than
10% outside the set measuring
range.
151.2
µµS/cm
13.5 min
method DEFAULT
Current time
The following possibilities exist for the display of
the current time:
Without "INJECT A" or program start
Current time after instrument switched on
With "INJECT A"
Current time after the last switching of injection
valve A to "INJECT" position (does not apply to
an "INJECT A" command within a program)
With program start
Current time from last program start