2.3 Connection of 733 IC Separation Center

732 IC Detector / 733 IC Separation Center

13

2.3

Connection of 733 IC Separation Center

2.3.1

2.733.0010/2.733.0X30 IC Separation Center

The instrument versions 2.733.0010 and 2.733.0030 of the IC Separa-

tion Center are operated with a 732 IC Detector whose standard

equipment also includes the 1.732.0100 Detector block. For the

metal-free 2.733.0130 instrument version the metal-free 1.732.0110

Detector block must be used. It is best to proceed as follows when

connecting the two instruments and the detector block:

1 Install detector block

Unscrew the four knurled screws 34

34 from the top rear panel

35

35 of the 733 IC Separation Center and remove rear panel

(see Fig. 5).

Position detector block from the back in the space provided

in the 733 IC Separation Center on the right and push fully to

the front (see Fig. 16).

Insert the cable permanently attached to the detector block in

opening 33

33 and the outlet capillary in opening 32

32 "Waste A" of

the rear panel 35

35.

Replace rear panel 35

35 and screw to the 733 IC Separation

Center using the four knurled screws 34

34.

2 Connect detector block

Plug the gray connecting cable permanently attached to the

detector block into connection 13

13 "Detector Block" of the 732

IC Detector and fasten to the instrument by tightening the

screws in the cable connector (see Fig. 6).

RUN COM

6.2125.090

Fig. 6: Connection 732 – 2.733.0010/2.733.0X30

Cable from

detector block