3 Operating tutorial

732 IC Detector / 733 IC Separation Center

58

10 Start 709 IC Pump

If you do not operate the 709 IC Pump by remote control

via the 732 IC Detector, the pump must be switched on

directly using the key (see 709 Instructions

for Use).

PUMP R/S

8

Press the key. The pump drive of the 709 IC

Pump is started.

151.2

µµS/cm 8.9 min

pump running 6.9 MPa

After the start of the 709 IC Pump, an automatic switch is

made to the status message for the pump. If this is running

properly, the message

"pump running"

and the current

measured pressure appear.

11 Condition IC system

Rinse IC system with eluent until the desired stability of the

baseline is achieved.
To assess the baseline stability, it is advantageous if the

measured conductivity is continuously recorded with a re-

corder or printer or (e.g. with «IC Metrodata for Win95»)

shown on the PC.
It normally takes 30

60 min until the IC system is ready for

analysis. If the eluent is changed, the establishment of the

ion exchanger equilibrium on the separating column may

take longer.

151.2

µµS/cm 53.2 min

pump running 6.9 MPa

Example: With the PRP-X100 IC Anion Column a stable

conductivity value of ca. 150

µ

S/cm is reached after around

45 min and scarcely changes after this time.