5.4 Diagnostic tests

732 IC Detector / 733 IC Separation Center

157

The test runs automatically and lasts ca. 3 min. During this time various measurement
results appear in the display, but these should be ignored. If the test discovers results
which lie outside the specified tolerance zones, the number of deviations found ap-
pears following completion of the test. If all results of the test are within the specified
tolerance zones, the following appears:

err: 0
DAC fine test ok

>diagnosis
>diag/external test

5.4.9

Initialize data memory

This diagnostic step can be used to write default values to the instru-

ment parameters using the keypad and thus switch the instrument to

the original condition. This measure is important with the following two

points:

The setting of certain instrument parameters such as the locking of

keys is possible only via RS232, i.e. with the aid of a PC. If such

instrument parameters are set and no PC is available to cancel the

settings, full use can not be made of the instrument.

In rare cases, it is possible that major interference signals such as line

spikes and lightning can have an adverse influence on the contents of

the data memory. If the contents of the data memory are undefined,

this may lead to a system crash.

With the initialization step, default values can be written into parts of the

data memory (EEPROM). Although the adjustment data are retained,

the initialization should be performed only if necessary as the stored

user data (etc.) are deleted.

Prepare instrument for diagnostic test (see section 5.4.2).

If need be, press <9> key repeatedly until

diagnosis
>diag/EEPROM initial.