Metrohm 732 IC Detector User Manual

Page 108

background image

4.5 Measurement parameters

732 IC Detector / 733 IC Separation Center

101

>PARAM
>PARAM/709 IC Pump

Settings for 709 IC Pump

This submenu appears only with "

SETUP/peripherals/operation with

709 = on

"

.

These settings are effective only if the 709 IC Pump is con-

nected to the 732 IC Detector and the remote control is acti-

vated with the key (see section 2.6.1).

>PARAM/709 IC Pump
flow 0.50 mL/min

0.01

5.00 mL/min

Flow rate
Flow rate of the 709 IC Pump.

>PARAM/709 IC Pump
Pmax 10.0 MPa

0.1

50 MPa

Maximum shutoff pressure
Maximum shutoff pressure for the automatic shutoff of

the pump drive of the 709 IC Pump to protect the

separating column against excessive pressure.
This limit value should be ca. 3 MPa (= 30 bar) above

the normal operating pressure of the separating column

used or correspond to the maximum admissible oper-

ating pressure of the separating column.
If the 709 Pump exceeds the specified limit value, the

pump drive is switched off within a pump cycle and

blocked, at the same time the bottom line of the display

of the 732 IC Detector shows the message

"

stopped:

YY-MM-DD HH:MM

"

with details of the time when the

pump was stopped. The LED

"

P

max

"

on the 709 IC

Pump flashes, the shutoff pressure can be requested

under

"

P

actual

"

.

A 709 IC Pump stopped in this manner is put back into

operation as follows:

Press key to switch off remote opera-

tion.

Press key to remove block. The LED

"

P

max

"

no longer flashes.

Press key to switch remote operation

back on.

>PARAM/709 IC Pump
Pmin 0.0 MPa

0.0

50 MPa

Minimum shutoff pressure
Minimum shutoff pressure for the automatic shutoff of

the pump drive of the 709 IC Pump to protect the IC

system when leaks occur or the eluent supply is inter-

rupted.
This limit value should be sufficiently far below the

normal operating pressure of the separating column

used. The entry of

"

0.0

"

switches the automatic shutoff

off.
If the pressure value of the 709 IC Pump is below the

specified limit value for a considerable length of time,

the pump drive of the pump is switched off and

blocked, at the same time the bottom line of the display