5.2 Maintenance and servicing

732 IC Detector / 733 IC Separation Center

137

corrosive chemicals, it may be necessary to shorten the interval be-

tween servicing.
The Metrohm service department is always willing to offer expert advice

on the maintenance and servicing of all Metrohm instruments.

5.2.2

Passivation

Passivation of the entire IC system (without column) by rinsing with

20...50 mL 0.2 mol/L HNO

3

is appropriate in the following cases:

in the installation

(only with non metal-free IC systems, see section 2.6.7)

if the separation system is changed

(only with non metal-free IC systems, see section 5.2.5)

if exceptional changes are observed in the

measurement properties of the cell

(with metal-free and non metal-free IC systems).

For passivation the separating column is removed from the 733 IC

Separation Center. The two capillaries 67

67 and 82

82 (see Fig. 16 and

Fig. 17) are directly connected to each other using the 6.2620.060

Coupling enclosed with the accessories.

5.2.3

Recycling

To keep the eluent consumption between injections to a minimum

when the system is at rest (e.g. overnight), the so-called recycling pro-

cedure can be used. In recycling the eluent exiting the outlet capillary of

the detector block is led back directly to the eluent supply vessel. The

IC system is thus quickly ready for new injections without a long condi-

tioning period.

The recycling procedure must not be used

in operation with the suppressor module,

with alkaline eluents,

with the 6.1010.000 IC Cation column METROSEP Cation 1-2.

5.2.4

Shutdown

If the 732/733 IC-System is shut down for a considerable length of time,

the entire IC system (without column and suppressor) must be rinsed

free from salt with methanol/water (1:4) to avoid crystallization of elu-

ent salts with the corresponding subsequent damage.
For rinsing the separating column is removed; the two capillaries 67

67

and 82

82 (see Fig. 16 and Fig. 17) are directly connected to each other

with the 6.2620.060 Coupling enclosed with the accessories. Rinse with

methanol/water (1:4) until the conductivity drops below 10

µ

S/cm.