5 Notes – Maintenance – Faults

732 IC Detector / 733 IC Separation Center

146

Error message

Meaning/Causes

Rectification

Err5

RS232 interface at 732 IC Detector

faulty.

Inform Metrohm service.

Err6

Clock faulty.

Inform Metrohm service.

Err7

LCD write/read error.

Inform Metrohm service.

Err8

AD converter faulty.

Inform Metrohm service.

ErrC

RS232 interface at 709 IC Pump

faulty.

Inform Metrohm service.

faulty method

#

Method

#

(0

9) can no longer be

recalled.

Re-enter method and store.

inject in inject program

The active inject program contains

an "

inject

" command. As this itself

triggers the program start, the

program will be executed endlessly.

Delete "

inject

" command in the

program.

IO error 50

Error at input 1 or output 1.

Inform Metrohm service.

IO error 51

Error at input 2 or output 2.

Inform Metrohm service.

IO error 52

Error at input 3 or output 3.

Inform Metrohm service.

IO error 53

Error at input 4 or output 4.

Inform Metrohm service.

IO error 54

Error at input 5 or output 5.

Inform Metrohm service.

IO error 55

Error at input 6 or output 6.

Inform Metrohm service.

IO error 56

Error at input 7 or output 7.

Inform Metrohm service.

IO error 57

Error at input 8 or output 8.

Inform Metrohm service.

IO error 58

Frequency less that lower limit

frequency of an RC element.

Inform Metrohm service.

IO error 59

Frequency greater that upper limit

frequency of an RC element.

Inform Metrohm service.

method memory full

The memory for the user-defined

methods is full.

Delete unused or seldom needed

methods.

missing 'end'

An “

end

” command is missing in the

active program.

Insert program step with “

Flag = end

in the active program. Without an “

end

command, the program continues to run

indefinitely until stopped by

R/S>.

missing 'return'

In the active loop program there is

no "

return

" command so that the

program runs only once.

Enter "

return

" command in the pro-

gram.

multiple 'end'

Active program contains several

"

end

" commands.

Delete superfluous "

end

" commands in

the program.

multiple 'return'

Active program contains several

"

return

" commands.

Delete superfluous "

return

" commands

in the program.

plot data overflow

Volume of data too great for graph-

ics plot.

(graphics plot aborted).

program step after 'end'

The active program contains addi-

tional program steps after an "

end

"

command which can no longer be

executed.

Delete program steps after "

end

" or

increase time for "

end

" command.

pump not responding

No connection to 709 IC Pump or

baud rate set wrongly.

Connect 709 IC Pump to 732 IC Detec-

tor, switch on pump and remote control

with or check baud rate of 709

and 732.

RAM error

Error in RAM test.

Inform Metrohm service.