6.1 RS232 interfaces

732 IC Detector / 733 IC Separation Center

177

6.1.7

Data transmission protocol

The two RS232 interfaces of the 732 IC Detectors are configured as

DTE (Data Terminal Equipment) with the following technical specifica-

tions:

Standard

Data interface in accordance with EIA standard

RS 232C (DIN 66020, page 1), transmission

parameters can be set under "

>CONFIG/RS

settings

", see section 4.4.2.

Control characters C

R

DEC 13

HEX 0D

L

F

DEC 10

HEX 0A

XON

DEC 17

HEX 11

XOFF

DEC 19

HEX 13

Max. line length

80 characters + C

R

L

F

Mode

Full duplex (simultaneous transmission and

receiving)
Restriction:

If data are first received by the interface, trans-

mission is not started until the receipt is at an

end.

Cable

For interconnection of the 732 IC Detector with

third-party devices, only a shielded data cable

may be used (e.g. Metrohm D.104.0201). The

cable shield must be perfectly earthed at both

devices (pay attention to current loops: always

use star earthing). Use only connectors with

sufficient shielding (e.g. Metrohm K.210.0001

with K.210.9004).

Cable length

max. ca. 15 m

Start

7 or 8 data bits

parity bit

1 or 2 stop bits