Page 113

background image

4 Operation

732 IC Detector / 733 IC Separation Center

106

4.6.4

key

ZERO OFF

4

The key is used to switch off the auto-zero function. The

auto-zero function can be triggered using the key in the

standby mode or when a program is running, but not in the edit mode.
When the green LED 66 in the key goes out, this shows that

the electronic background compensation is no longer active.

4.6.5

key

MARK

0

The key is used to trigger a marking signal. In addition to the

measuring signal, a signal of ca. 10% of the full-scale range is output-

ted briefly at the analog output. This allows, e.g. the start of a chroma-

togram to be marked during recording on a recorder.
Marking can be triggered with the key in the standby mode

or when a program is running, but not in the edit mode.

4.6.6

key

PUMP R/S

8

The

the 709 IC Pump. A requirement for this is that the 709 IC Pump is

connected to the 732 IC Detector and the remote control is switched on

at the pump with the key (see section 2.6.1).
With the key the pump drive of the 709 IC Pump can be

switched on or off in the standby mode or when a program is running,

but not in the edit mode. When the pump drive is switched on, the red

LED above the key on the pump lights up.
The settings of the pump parameters on the 732 IC Detector 732 are

described under the key in section 4.4.2, the pump status

messages in the standby mode in section 4.3.2.