Metrohm 732 IC Detector User Manual
Page 92
4.4 Basic settings
732 IC Detector / 733 IC Separation Center
85
>SETUP
>SETUP/input assign
Assignment of the remote input lines
The 732 IC Detector has 4 programmable remote input lines
which can be used to trigger a total of 15 different instrument
functions (for details, see section 6.2.1). The default assign-
ment of the individual functions to the status of the 4 input
lines 2
…
5 (defined by the corresponding decimal code
1
…
15) can be changed if need be. Several functions can
thus also be assigned the same status.
>SETUP/input assign
PROG R/S 1
1
…
15
Start/stop program
An active remote program is started, any programs
running are stopped.
>SETUP/input assign
PUMP R/S 2
1
…
15
Start/stop 709 IC Pump
Equivalent to pressing the
>SETUP/input assign
FILL A 4
1
…
15
Trigger "FILL" at valve A
Equivalent to pressing the
Separation Center.
>SETUP/input assign
INJECT A 8
1
…
15
Trigger "INJECT" at valve A
Equivalent to pressing the
IC Separation Center.
>SETUP/input assign
FILL B/STEP 3
1
…
15
Trigger "FILL" at valve B or "STEP" on the
suppressor module
Equivalent to pressing the
>SETUP/input assign
INJECT B 12
1
…
15
Trigger "INJECT" at valve B
Equivalent to pressing the
IC Separation Center.
>SETUP/input assign
REPORT 5
1
…
15
Report output
Equivalent to pressing the
>SETUP/input assign
ZERO 6
1
…
15
Switch on auto-zero
Equivalent to pressing the
>SETUP/input assign
MARK 7
1
…
15
Switch on marking
Equivalent to pressing the key.