130 method, 4 operation – Metrohm 732 IC Detector User Manual

Page 137

background image

4 Operation

732 IC Detector / 733 IC Separation Center

130

Method

METHOD
method name An.Dual
date: 1995-10-16 14:30:39

PARAM
>PARAM/detector
range: 100

µ

S/cm

full scale: 50.00

µ

S/cm

temp.coeff.: 2.5 %/°C
>PARAM/analog output
polarity: +
offset: 0 %fs
damping: off
>PARAM/Plot
auto start: off
time interval 1.0 s
time scale 10.0 mm/min
time scale label: rel
stop time off min
left: -5.000

µ

S/cm

right: 45.00

µ

S/cm

>PARAM/709 IC Pump
flow 0.80 mL/min
Pmax 7.0 MPa
Pmin 1.0 MPa
flow corr. 1.00

PROGRAM
progr.type: cycle
number of cycles 1
9.0 range: 100

µ

S/cm

full scale:5.00

µ

S/cm

mark: on
14.7 range: 100

µ

S/cm

full scale:50.0

µ

S/cm

mark: on
19.9 zero: on
49.9 valve A: fill
remote 0*******
pump R/S: off
50.0 flag: end
--------------

Switching of full-scale range to 5

µ

S/cm

Triggering of a marking signal
Switching of full-scale range to 50

µ

S/cm

Triggering of a marking signal
Triggering of the auto-zero function
Switching of valve A to "FILL"
Switching off of 752 Pump Unit
Switching off of 709 IC Pump
Program end

The program of the type "inject" is started each time injection valve A is

switched to the "INJECT" position. For the determination of nitrate and

phosphate, the full-scale range is briefly narrowed to 5

µ

S/cm and this

range marked by marking signals. After 20.0 min the IC system is ready

for the next injection. If there is no new program start within 50 min after

the last injection, valve A is automatically switched to the "FILL" position,

the 709 IC Pump and 752 Pump Unit are stopped and the program

ended.