4 Operation

732 IC Detector / 733 IC Separation Center

88

4.4.2

Configuration, key

7

CONFIG

The key is used for the entry of basic settings of general

applicability which are also retained when the instrument is switched

off. The key opens the following main menu:

CONFIG
>CONFIG/detector

Settings for the conductivity detector

CONFIG
>CONFIG/printer

Settings for external printer

CONFIG
>CONFIG/print meas.value

Settings for measured value printout

CONFIG
>CONFIG/auxiliaries

Various general instrument settings

CONFIG
>CONFIG/RS settings

Settings for interface "RS 232"

CONFIG
>CONFIG/RS settings 709

Settings for interface "709 IC Pump" (RS232)

CONFIG
>CONFIG/733 IC Sep.Cent.

Settings for 733 IC Separation Center

Repeated pressing of the key selects the submenus in

turn. The individual inquiries of a submenu are accessed using the

key and exited with the key. The following listings

show all dialog items which appear under . The values

shown in the display are the default values, the possible entry values or

ranges are shown below the display.

>CONFIG
>CONFIG/detector

Settings for the conductivity detector

>CONFIG/detector
thermostat: 35 °C

25, 30, 35, 40, 45 °C, off

Thermostatting of the conductivity detec-
tor
Setting of the operating temperature of the conductivity

cell.
At constant ambient temperature the heating built into

the 732 IC Detector regulates the selected measuring

cell temperature to an accuracy of ± 0.01 °C and thus

establishes the precondition for highly sensitive deter-

minations. It normally takes 30...60 min after the instru-

ment has been switched on until this temperature sta-

bility is attained.