Metrohm 732 IC Detector User Manual
Page 95
![background image](/manuals/586520/95/background.png)
4 Operation
732 IC Detector / 733 IC Separation Center
88
4.4.2
Configuration,
7
CONFIG
The
applicability which are also retained when the instrument is switched
off. The key opens the following main menu:
CONFIG
>CONFIG/detector
Settings for the conductivity detector
CONFIG
>CONFIG/printer
Settings for external printer
CONFIG
>CONFIG/print meas.value
Settings for measured value printout
CONFIG
>CONFIG/auxiliaries
Various general instrument settings
CONFIG
>CONFIG/RS settings
Settings for interface "RS 232"
CONFIG
>CONFIG/RS settings 709
Settings for interface "709 IC Pump" (RS232)
CONFIG
>CONFIG/733 IC Sep.Cent.
Settings for 733 IC Separation Center
Repeated pressing of the
turn. The individual inquiries of a submenu are accessed using the
show all dialog items which appear under
shown in the display are the default values, the possible entry values or
ranges are shown below the display.
>CONFIG
>CONFIG/detector
Settings for the conductivity detector
>CONFIG/detector
thermostat: 35 °C
25, 30, 35, 40, 45 °C, off
Thermostatting of the conductivity detec-
tor
Setting of the operating temperature of the conductivity
cell.
At constant ambient temperature the heating built into
the 732 IC Detector regulates the selected measuring
cell temperature to an accuracy of ± 0.01 °C and thus
establishes the precondition for highly sensitive deter-
minations. It normally takes 30...60 min after the instru-
ment has been switched on until this temperature sta-
bility is attained.