5.3 Faults and malfunctions

732 IC Detector / 733 IC Separation Center

145

5.3

Faults and malfunctions

5.3.1

Error messages

If any type of malfunction occurs during operation of the 732 IC Detec-

tor, this is shown by an error message which appear in the top line of

display 11 and (if switched on under ) by three beeps. To

return to the instrument standby mode, the error messages must be

confirmed with the key.

You will find further details of the error messages, their possible causes

and the procedure for their rectification in the following table, which is

arranged alphabetically. If the errors can not be rectified, please follow

the procedure described in section 5.4.

Error message

Meaning/Causes

Rectification

adjustment error

Instrument adjustment invalid.

Inform Metrohm service.

all methods faulty

The stored methods can no longer

be recalled.

Re-enter methods and store. If this error

occurs frequently, inform Metrohm

service.

connect 709 IC Pump

The active program contains a

command for the 709 IC Pump,

which can not be addressed at

present.

Check connection to 709 IC Pump,

switch on pump and remote operation

with .

connect 733 Sep.Center

The active program contains a

command for the 733 IC Separation

Center, which can not be addressed

at present.

Check connection to 733 IC Separation

Center.

DAC coarse error

#

Error in internal hardware test.

Inform Metrohm service.

DAC fine error

#

Error in internal hardware test.

Inform Metrohm service.

EEPROM error 233

Error on storage of the EEPROM

block "Program".

Re-enter program.

EEPROM error 234

Error on storage of the EEPROM

block "Parameters".

Re-enter parameters.

EEPROM error 235

Error on storage of the EEPROM

block "Method".

Re-enter method (program and pa-

rameters).

EEPROM error 236

Error on storage of the EEPROM

block "Event".

Re-enter events.

EEPROM error 237

Error on storage of the EEPROM

block "Config".

Re-enter configuration.

EEPROM error 238

Error on storage of the EEPROM

block "Device number ".

Re-enter device number.

EEPROM error 239

Error on storage of the EEPROM

block "Adjustment".

Inform Metrohm service.

EEPROM error 240

Error on storage of the EEPROM

blocks.

Re-enter all values.

Err1

Program checksum wrong.

Inform Metrohm service.

Err3

Instrument adjustment faulty.

. If new instrument adjustment

necessary, inform Metrohm service.

Err4

Timer interrupt for multitasking

missing.

Inform Metrohm service.