4.5 Measurement parameters

732 IC Detector / 733 IC Separation Center

99

>PARAM/detector
temp.coeff.: 2.5 %/°C

1.5, 2.5 %/°C

Temperature coefficient
The temperature coefficient is used to take into account

the temperature dependence of the conductivity meas-

urement. Two values can be selected:

1.5 %/°C

Temperature coefficient for cations

2.5 %/°C

Temperature coefficient for anions

When the thermostatting of the measuring cell is

switched on (thermostat display 77 lit up), the conductiv-

ity

κκ

T

measured at the set operating temperature

(25...45 °C) is automatically converted to the conductiv-

ity

κκ

20

at the reference temperature 20 °C.

>PARAM
>PARAM/analog output

Settings for the analog output
The output signal at the two analog outputs 11

11 (0

1 V) and

12

12 (0

10 mV) is determined by the parameter

"

full scale

"

(see above).

>PARAM/analog output
polarity: +

+,–

Polarity of the output signal
Selection of whether a positive or negative full-scale

range should be outputted

+

Positive polarity

(0

+FS)

Negative polarity (0

FS)

>PARAM/analog output
offset: 0 %fs

0, 10, 50 %fs

Zero point offset
Offset of the zero point of the conductivity.

0 %fs

No zero point offset.

10 %fs

Offset by 10% of the full-scale range;

at the analog output the signal

–0.1 FS

+0.9 FS (+) or

0.1 FS

–0.9 FS (–) is outputted.

50 %fs

Offset by 50% of the full-scale range;

at the analog output the signal

–0.5 FS

+0.5 FS (+) or

0.5 FS

–0.5 FS (–) is outputted.

>PARAM/analog output
damping: off

on, off

Damping

on

Damping to reduce disturbing pulsations

or a high noise level. Standard solutions

and samples must both be recorded with

damping as this can influence the peak

heights.

off

Damping switched off (normal condition

with pulsation-free and constant flow

pumps).