beautypg.com

Metrohm 732 IC Detector User Manual

Page 147

background image

5 Notes – Maintenance – Faults

732 IC Detector / 733 IC Separation Center

140

5.2.6

Regeneration of the suppressor

Regenerating a suppressor operating at

reduced capacity
If the suppressor units are exposed to certain heavy metals (e.g. iron)

or organic contaminants for long periods of time, these can no longer

be completely removed by the regeneration solution normally used (20

mmol/L H

2

SO

4

). This diminishes the capacity of the suppressor units,

which, in milder cases, results in a reduced sensitivity to phosphates

and, in severe cases, in a strong increase in the baseline. If such ca-

pacity problems occur at one or several positions, the suppressor units

must be treated as follows:

1 Disconnect suppressor from IC system

Disconnect suppressor from separating column and detector.

2 Regenerate suppressor

Rinse each suppressor unit for about 15 min with one of the

following solutions:

Contamination with heavy metals

1 mol/L H

2

SO

4

Contamination with organic cationic complexing

agents

0.1 mol/L H

2

SO

4

/ 0.1 mol/L oxalic acid / acetone 5%

Severe contamination with organic substances

0.2 mol/L H

2

SO

4

/ acetone

20%

The 6.1826.050 pump tubing is made of PVC and must

not be used for rinsing with solutions which contain

acetone. In such cases, rinse with different pump tubing

or a different pump.

3 Connect suppressor to IC system

Reconnect suppressor to the IC system. If capacity problems

persist, replace the suppressor rotor (see section 5.2.8).

Regenerating the suppressor in case of

high counterpressure
If excessive counterpressure is observed in one or several suppressor

units, treat the units as follows:

1 Disconnect suppressor from IC system

Disconnect suppressor from separating column and detector.