5 Notes – Maintenance – Faults

732 IC Detector / 733 IC Separation Center

158

Press key to open the following diagnostic menu:

>diag/EEPROM initial.
block: USER

ь

ь ь

ь ь

ь ь

ь ь

ь ь

ь

>diag/EEPROM initial.
block: CONFIG

ь

ь

>diag/EEPROM initial.
block: INFO

ь

ь

>diag/EEPROM initial.
block: SETUP

ь

ь

>diag/EEPROM initial.
block: ASSEMBLY

ь

ь

>diag/EEPROM initial.
block: EVENT

ь

ь

>diag/EEPROM initial.
block: METHOD

ь

ь

>diag/EEPROM initial.
block: ALL

ь

ь ь

ь ь

ь ь

ь ь

ь ь

ь ь

ь ь

ь

1)

Set config parameters with default values.

Ã

Set info parameters with default values.

Ã

Set setup parameters with default values.

Ã

Set assembly parameters with default values.

Ã

Set event parameters with default values.

Ã

Delete method memory.

Ã

Delete device number.

Ã

Set instrument adjustment with default values.

Ã

Pressing the key repeatedly until the desired

initialization alternative appears (such as “USER“)

>diag/EEPROM initial.
block: USER

After the initialization has run automatically, the instrument quits the diagnostic menu
and runs a power on reset.

1)

Care:

This initialization alternative is reserved for the service engineer!