Output of individual measured values – Metrohm 732 IC Detector User Manual

Page 127

background image

4 Operation

732 IC Detector / 733 IC Separation Center

120

Identification

"

ZEROREF

"

Reference value for the auto-zero function (measured value which is

compensated).

Identification

"

OVERLOAD

"

The measured value is 50

80% outside the selected full-scale

range.

Identification

"

BACKGROUND

"

The measured value is more than 80% outside the selected full-scale

range. The absolute conductivity (background) is outputted in place

of the auto-zero value.

Output of individual measured values

For the output of individual measured values,

"

>CONFIG/print

meas.value

=

immed.

"

must be selected (see section 4.4.2). Each

time the key is pressed, a measured value will be out-

putted via the RS232 interface.

Example:

Individual measured value output with print header output

(with date and time), run time, measured value

>CONFIG/printer
id.1
id.2
print header: always
date&time: on
send to: Seiko
>CONFIG/print meas.value
print crit.: immed.
date&time: off

732 IC Detector 01104 732.0012

Print header

date 1995-10-09 time 13:24:58

Date and time

12.3 138.9 uS/cm

Run time, measured value

==============

732 IC Detector 01104 732.0012

Print header

date 1995-10-09 time 13:25:37

Date and time

139.1 uS/cm ZEROREF

Auto-zero reference value

13.0 +0.0 uS/cm

Run time, auto-zero value

==============