Output of individual measured values – Metrohm 732 IC Detector User Manual
Page 127
![background image](/manuals/586520/127/background.png)
4 Operation
732 IC Detector / 733 IC Separation Center
120
•
Identification
"
ZEROREF
"
Reference value for the auto-zero function (measured value which is
compensated).
•
Identification
"
OVERLOAD
"
The measured value is 50
…
80% outside the selected full-scale
range.
•
Identification
"
BACKGROUND
"
The measured value is more than 80% outside the selected full-scale
range. The absolute conductivity (background) is outputted in place
of the auto-zero value.
Output of individual measured values
•
For the output of individual measured values,
"
>CONFIG/print
meas.value
=
immed.
"
must be selected (see section 4.4.2). Each
time the
putted via the RS232 interface.
•
Example:
Individual measured value output with print header output
(with date and time), run time, measured value
>CONFIG/printer
id.1
id.2
print header: always
date&time: on
send to: Seiko
>CONFIG/print meas.value
print crit.: immed.
date&time: off
732 IC Detector 01104 732.0012
Print header
date 1995-10-09 time 13:24:58
Date and time
12.3 138.9 uS/cm
Run time, measured value
==============
732 IC Detector 01104 732.0012
Print header
date 1995-10-09 time 13:25:37
Date and time
139.1 uS/cm ZEROREF
Auto-zero reference value
13.0 +0.0 uS/cm
Run time, auto-zero value
==============