4 Operation

732 IC Detector / 733 IC Separation Center

100

>PARAM
>PARAM/plot

Settings for graphics plot
These settings are effective with the graphics plot on an

external printer triggered using the key (see sec-

tion 4.8.2).

>PARAM/plot
auto start: off

on, off

Automatic start

on

Automatic start of the graphics plot on every

"INJECT A".

off

No automatic start of the graphics plot.

>PARAM/plot
time interval 1.0 s

0.4

99999 s

Time interval
Time interval after which a new measured value is plot-

ted.

>PARAM/plot
time scale 10.0 mm/min

0.1, 0.2, 0.5, 1, 2, 5,

10, 20, 60, 120 mm/min

Time scale (paper speed)
The entry value corresponds to the paper feed of the

printer in mm/min (the effective paper feed depends on

the printer used).
If values other than those specified are entered, these

will be automatically rounded to the next preset value.

>PARAM/plot
time scale label: rel

rel, abs

Labeling of the time axis
The time axis is labeled at every main division.

abs

Labeling with absolute (current) time

(e.g.

"

08:34

"

).

rel

Labeling with relative time beginning with

the start of the curve printout

(e.g.

"

2m40s

"

for 2 min 40 s).

>PARAM/plot
stop time off min

1

999 min, off

Stop time for graphics plot
Time until graphics plot is stopped.

off

No time limit (press ).

>PARAM/plot
left: 0.000

µµS/cm

-2000

2000

µ

S/cm

-2000

2000 mS/cm

-2

2 S/cm

Left boundary of conductivity axis
Left limit value for graphics plot of the conductivity on an

external printer.
The unit (

µ

S/cm, mS/cm, S/cm) can be selected with

the key.