4 Operation
732 IC Detector / 733 IC Separation Center
100
>PARAM
>PARAM/plot
Settings for graphics plot
These settings are effective with the graphics plot on an
external printer triggered using the
tion 4.8.2).
>PARAM/plot
auto start: off
on, off
Automatic start
on
Automatic start of the graphics plot on every
"INJECT A".
off
No automatic start of the graphics plot.
>PARAM/plot
time interval 1.0 s
0.4
…
99999 s
Time interval
Time interval after which a new measured value is plot-
ted.
>PARAM/plot
time scale 10.0 mm/min
0.1, 0.2, 0.5, 1, 2, 5,
10, 20, 60, 120 mm/min
Time scale (paper speed)
The entry value corresponds to the paper feed of the
printer in mm/min (the effective paper feed depends on
the printer used).
If values other than those specified are entered, these
will be automatically rounded to the next preset value.
>PARAM/plot
time scale label: rel
rel, abs
Labeling of the time axis
The time axis is labeled at every main division.
abs
Labeling with absolute (current) time
(e.g.
"
08:34
"
).
rel
Labeling with relative time beginning with
the start of the curve printout
(e.g.
"
2m40s
"
for 2 min 40 s).
>PARAM/plot
stop time off min
1
…
999 min, off
Stop time for graphics plot
Time until graphics plot is stopped.
off
No time limit (press
>PARAM/plot
left: 0.000
µµS/cm
-2000
…
2000
µ
S/cm
-2000
…
2000 mS/cm
-2
…
2 S/cm
Left boundary of conductivity axis
Left limit value for graphics plot of the conductivity on an
external printer.
The unit (
µ
S/cm, mS/cm, S/cm) can be selected with
the
>PARAM/plot
right: 10.00
µµS/cm
-2000
…
2000
µ
S/cm
-2000
…
2000 mS/cm
-2
…
2 S/cm
Right boundary of the conductivity axis
Right limit value for graphics plot of the conductivity on
an external printer.
The unit (
µ
S/cm, mS/cm, S/cm) can be selected with
the