4 Operation

732 IC Detector / 733 IC Separation Center

108

â

Program steps 1

k

â

return

â

Program steps m

n

â

end

No. of cycles

remote

Single program with start via remote

A remote program is started externally via a

remote input line (see section 6.2.1) and

executed once only.

inject

Single program with start via "INJECT"

An inject program is started each time injec-

tion valve A is switched to the "INJECT" posi-

tion and executed once only.

>PROGRAM/parameters
number of cycles 1

1

999

Number of cycles

This inquiry appears only with "

PROGRAM/parameters/type =

cycle

"

.

Number of executions of loop programs.

This parameter is identical to that of the same name in

the submenu

"

>CONFIG/auxiliaries

"

of the

FIG> key (see section 4.4.2). As access to the

key is blocked while a program is

running, the number of cycles can be changed only

there.

>PROGRAM/parameters
status: disabled

enabled, disabled, test

Program status
The three following program statuses are possible:

enabled

Program enabled

Irrespective of the program type, a pro-

gram can be started only when it is en-

abled. This status is shown by the lighting

up of the green LED 55 in the

R/S> key. If a method with program

steps is loaded, the program status is

automatically set to "

enabled

".

disabled

Program disabled

A disabled program can not be started. In

this status, the green LED 55 in the

key remains off. When the

instrument is switched on, the program

status is always set to

"

disabled

"

.