4 Operation
732 IC Detector / 733 IC Separation Center
108
â
Program steps 1
…
k
â
return
â
Program steps m
…
n
â
end
No. of cycles
remote
Single program with start via remote
A remote program is started externally via a
remote input line (see section 6.2.1) and
executed once only.
inject
Single program with start via "INJECT"
An inject program is started each time injec-
tion valve A is switched to the "INJECT" posi-
tion and executed once only.
>PROGRAM/parameters
number of cycles 1
1
…
999
Number of cycles
This inquiry appears only with "
PROGRAM/parameters/type =
cycle
"
.
Number of executions of loop programs.
This parameter is identical to that of the same name in
the submenu
"
>CONFIG/auxiliaries
"
of the
FIG> key (see section 4.4.2). As access to the
running, the number of cycles can be changed only
there.
>PROGRAM/parameters
status: disabled
enabled, disabled, test
Program status
The three following program statuses are possible:
enabled
Program enabled
Irrespective of the program type, a pro-
gram can be started only when it is en-
abled. This status is shown by the lighting
up of the green LED 55 in the R/S> key. If a method with program steps is loaded, the program status is automatically set to " enabled ". disabled Program disabled A disabled program can not be started. In this status, the green LED 55 in the instrument is switched on, the program status is always set to " disabled " .