3.3 Putting into operation

732 IC Detector / 733 IC Separation Center

57

>PARAM/709 IC Pump
Pmax 10.0 MPa

The maximum admissible pressure for the PRP-X100 IC

Anion Column is 34 MPa, the normal operating pressure is

2 mL/min at ca. 7 MPa. A value of 10 MPa is thus entered

as the maximum shutoff pressure.

ENTER

Confirm the set maximum shutoff pressure with the

key.

>PARAM/709 IC Pump
Pmin 0.0 MPa

The parameter

"

Pmin

"

denotes the minimum shutoff

pressure for the 709 IC Pump. If the pressure drops below

this value for a considerable length of time (e.g. owing to

leaks or if the eluent supply is interrupted), the pump is

automatically switched off.
Use the numeric keys to enter here the desired value. This

limit value should be sufficiently far below the normal oper-

ating pressure of the separating column used (1 MPa =

10 bar).

>PARAM/709 IC Pump
Pmin 1.0 MPa

A value of 1 MPa can be entered as the minimum shutoff

pressure for the PRP-X100 IC Anion Column.

ENTER

Confirm the set minimum shutoff pressure using the

key.

>PARAM/709 IC Pump
flow corr. 1.00

The parameter

"flow corr."

denotes the correction

factor for the flow of the 709 IC Pump. This correction

factor is used to match the displayed flow to the actual flow.
If you are interested in the exact display of the flow, this

correction factor must be determined by measurement of

the actual flow and then entered here (see section 3.3 of

the 709 Instructions for Use).

ENTER

Confirm the inputted flow correction with the

key. As this was the last inquiry of the last submenu of the

key, an automatic switch is made to the

standby mode.

151.2

µµS/cm 8.5 min

method DEFAULT

The 732 IC Detector is again in the standby mode and the

display shows the status messages.