beautypg.com

Metrohm 732 IC Detector User Manual

Page 197

background image

7 Appendix

732 IC Detector / 733 IC Separation Center

190

Conductivity detector

Construction

Thermostatted conductivity detector with 2 ring-

shaped steel electrodes

Measurement principle

Alternating current measurement with following

cell supply:

Meas. Range

Amplitude

Frequency

100

µ

S/cm

1 V peak

714 Hz

200

µ

S/cm

1 V peak

5000 Hz

500

µ

S/cm

1 V peak

5000 Hz

1 mS/cm

1 V peak

5000 Hz

2 mS/cm

0.5 V peak

5000 Hz

5 mS/cm

0.2 V peak

5000 Hz

10 mS/cm

0.1 V peak

5000 Hz

Cell volume

1.5

µ

L

Cell constant

16.7 /cm; adjustable 13.0

21.0 /cm

Maximum back pressure

for measuring cell

5.0 MPa (50 bar)

Thermostatting

Connectable dynamic control to adjustable

operating temperature, indicated by LED

Operating temperature

Adjustable in steps of 5°C from 25

45°C

Max. temperature deviation

±

1.5°C

Temperature stability

0.01°C at constant ambient temperature

Parts and controls

Display

LCD, 2 lines, each of 24 characters

(character height 5 mm)

Indicators

LED for auto-zero display

LED for display of program status

LED for overload display

LED for display of block heating

Keypad

Chemically resistant membrane keypad made of

polyester with function and numeric keys

Mains switch

At rear of instrument

Mains connection

Voltage

115 V: 100...120 V

±

10 %

230 V: 220...240 V

±

10 %

Switching with mains voltage selector in fuse

holder (see section 2.4.1)

Frequency

50...60 Hz

Power consumption

70 VA

Fuse

5 mm Ø, 20 mm length

100

120 V: 0.63 A (slow-blow)

220

240 V: 0.315 A (slow-blow)