4 Operation
732 IC Detector / 733 IC Separation Center
84
4.4
Basic settings
4.4.1
Setup
Switch on +
7
CONFIG
Under "Setup", various basic settings of the 732 IC
Detector are combined which normally seldom
need to be changed. They are accessible only
when the
time as the instrument is switched on. The
"SETUP" part which then appears comprises the
following main menu:
SETUP
>SETUP/output
Setting of the remote output lines
SETUP
>SETUP/input assign
Assignment of the remote input lines
SETUP
>SETUP/graphics
Setting of general graphics parameters
SETUP
>SETUP/peripherals
Setting of the peripherals 733 and 709
Repeated pressing of the
turn. The individual inquiries of a submenu are accessed using the
show all dialog items which appear under "SETUP". The values shown
in the displays are the default values, the possible entry values or
ranges are shown below the display.
>SETUP
>SETUP/output
Setting of the remote output lines
For further details on the remote interface, see section 6.2.
>SETUP/output
remote 00000000
0,1
Set the remote output lines 1…8
Sets the basic setting for the remote output lines when
the instrument is switched on.
Each of the remote output lines 1…8 can be set manu-
ally from left to right in turn:
0
Line off, inactive (open)
1
Line on, active (0 V)
Each time the instrument is switched on, the remote
output lines are reset in accordance with the parame-
ters entered here.