7 precolumns, General information on precolumns, Precolumns with twin cartridge holder – Metrohm 732 IC Detector User Manual

Page 35: Precolumns with cartridge head, Ic anion precolumn supersep

background image

2 Installation

732 IC Detector / 733 IC Separation Center

28

2.7

Precolumns

2.7.1

General information on precolumns

The use of easily exchangeable precolumns protects the separating

columns and prolongs their lifetime. The precolumns available from

Metrohm (see section 7.3.2) are either real precolumns or precolumn

cartridges, which are used together with the 6.2821.050 Twin cartridge

holder or a 6.2821.040 Cartridge head.

New IC precolumns are normally filled with solution and sealed at

both ends. Before the precolumn is installed in the system, it must be

ensured that this solution is freely miscible with the eluent used

(check manufacturer's specifications).

2.7.2

Precolumns with twin cartridge holder

The precolumn cartridges are installed in the 6.2821.050 Twin cartridge

holder as follows (see Fig. 15):

1 Install cartridge

Insert inlet capillary 75

75 with end piece for precolumn cartridge

in Manufit housing 62

62.

Insert outlet capillary 72

72 with end piece for precolumn car-

tridge in Manufit pressure screw 57

57.

Remove end caps from the precolumn cartridge 74

74 (the steel

mesh 73

73 and gaskets 59

59 are already inserted in the car-

tridge).

Fit the two capillary end pieces onto the precolumn cartridge

74

74 (comply with the flow direction if specified on the column).

Screw Manufit pressure screw 57

57 and Manufit housing 62

62

firmly together.

2 Connect precolumn

Fit connectors to inlet capillary 75

75 of the assembled precol-

umn (see section 2.5).

Connect inlet capillary 75

75 either using 6.2620.060 Coupling to

the column connection capillary 67

67 mounted on the injection

valve (see section 2.6.4) or directly to connection "4“ of injec-

tion valve A or B.

Shorten outlet capillary 72

72 of the precolumn to ca. 5 cm and

mount the connectors (see section 2.5).

3 Rinse the precolumn

Place a beaker beneath outlet capillary 72

72.

Switch on 709 IC Pump, rinse precolumn for ca. 10 min with

eluent and then switch off 709 IC Pump.