Metrohm 732 IC Detector User Manual

Page 45

background image

2 Installation

732 IC Detector / 733 IC Separation Center

38

1

3

2

89

89

92

92

93

93

94

94

90

90

91

91

Fig. 18:

Connections at suppressor module

Screw outlet capillary 94

94 to coupling 96

96 using a 6.2744.010

Compression fitting.

Screw inlet capillary 82

82 of detector block 81

81 to other end of

coupling 96

96.

6 Fix connection suppressor – detector block

Insert one of the column holders 84

84 (6.2027.030, 6.2027.040

or 6.2027.050) in mounting rail 83

83 and fasten coupling 96

96 in

the column holder.

7 Prepare 752 Pump Unit

Take two tubing cartridges out of the holder of the 752 Pump

Unit.

Lay a length of 6.1826.050 Pump tubing in each of the two

tubing cartridges and reinsert these in the holder without

kinking the pump tubing.

Regulate contact pressure on pump tubing in accordance

with the instructions printed on the pump.

Pump tubing is consumable material with a lifetime which depends on

the contact pressure. For this reason, raise the tubing cartridges

completely by loosening the clip on the right if the pump is switched

off for some considerable time (the set contact pressure remains

unchanged).

8 Suppressor connection 2: H

2

SO

4

Loosen rotary nipple screwed onto the interior side of con-

nection 27

27. Pull inlet capillary 90

90 marked with "H

2

SO

4

" (see

Fig. 17 and Fig. 18) by hand out of the opening of connection

27

27 as far as required. Retighten nipple on the interior side of

connection 27

27 to fix inlet capillary 90

90.

89

89

Suppressor inlet

capillary for eluent

90

90

Suppressor inlet

capillary for H

2

SO

4

91

91

Suppressor outlet

capillary for H

2

SO

4

92

92

Suppressor outlet

capillary for H

2

O

93

93

Suppressor inlet

capillary for H

2

O

94

94

Suppressor outlet

capillary for eluent

Eluent

H

2

O

Detector

Waste

Waste

H

2

SO

4