Metrohm 732 IC Detector User Manual

Page 131

background image

4 Operation

732 IC Detector / 733 IC Separation Center

124

4.8.3

key

REPORT

2

The key is used for the output of reports on an external

printer or PC connected to the RS232 interface.
The key opens the following inquiry:

XXXXXX

µµS/cm XXXX min

report: all

all, config, event,

method, method list,

param, program, pump

Selection of the report

Selection of the report which should be outputted:

all

All reports in the order "

config,

event,

method,

method list,

pump

"

config

Configuration report (parameters of the

key)

event

Event report (parameters of the

key)

method

Method report (comprises method name,

parameter and program report)

method

Method list (details of stored

list

methods)

param

Parameter report (parameters of the

key)

program

Program report (parameters of the

key)

pump

Report of the pump parameters

Example of "config" report

Configuration for 732 IC Detector with attached Seiko printer and 733.0010

IC Separation Center with 1 injection valve

CONFIG
>CONFIG/detector
thermostat: 35 °C
'zero' unit:

µ

S/cm

cell constant 16.7 /cm
>CONFIG/printer
id.1 PRP-X100
id.2 drinking water
print header: once
date&time: on
send to: Seiko
>CONFIG/print meas.value
print crit.: time
time interval 300.0 s
stop time 60 min
date&time: off
>CONFIG/auxiliaries
run number 0
number of cycles 1
>CONFIG/aux/event
enabled in program: off
date 1995-10-10
time 11:38:20
dialog: english
device label
program 732.0012

>CONFIG/aux/beeper
status: on
repeat time 60 s
>CONFIG/RS settings
baud rate: 9600
data bit: 8
stop bit: 1
parity: none
handshake: HWs
control via RS: on
>CONFIG/RS settings 709
baud rate: 9600
>CONFIG/733 IC Sep.Cent.
valve A
control: no restriction
>CONFIG/733/valve A
status inject
trigger: inject
--------------