beautypg.com

127 method – Metrohm 732 IC Detector User Manual

Page 134

background image

4.9 Examples of methods

732 IC Detector / 733 IC Separation Center

127

Method

METHOD
method name Cat.1-2
date: 1995-10-16 10:57:42

PARAM
>PARAM/detector
range: 1.00 mS/cm
full scale: 5.00

µ

S/cm

temp.coeff.: 1.5 %/°C
>PARAM/analog output
polarity: -
offset: 0 %fs
damping: off
>PARAM/plot
auto start: off
time interval 1.0 s
time scale 10.0 mm/min
time scale label: rel
stop time off min
left: 1.000

µ

S/cm

right: -4.000

µ

S/cm

>PARAM/709 IC Pump
flow 1.00 mL/min
Pmax 10.0 MPa
Pmin 1.0 MPa
flow corr. 1.00

PROGRAM
progr.type: inject
11.9 zero: on
12.0 flag: end
--------------

Triggering of auto-zero function
Program end

The program of the type "inject" is started each time injection valve A is

switched to the "INJECT" position. After 12.0 min, the IC system is ready

for the next injection.