6 Interfaces

732 IC Detector / 733 IC Separation Center

166

6.1.6

Remote control commands

The remote control tree can be divided into the following main

branches:

&

Root

Ã

C

onfig

Instrument configuration

Ã

M

ethods

Methods

Ã

P

arameter

Parameters

Ã

Pr

ogram

Program

Ã

S

electEvent

Events

Ã

Z

ero

On/off switching of auto-zero

Ã

Ma

rker

Triggering marking signal

Ã

Pl

ot

Start/stop graphics plot

Ã

Pri

nt

Start/stop measured value printout

Ã

I

nfo

Instrument information

Ã

Set

up

Settings of operating mode

Ã

A

ssembly

Data of the assemblies

À

D

iagnosis

Diagnostic tests

The following table lists all objects of the remote control tree. For the

unambiguous designation of the objects, the boldface characters suf-

fice. The meaning of the individual object is described here only in brief,

for more detailed information please refer to section 4. The default val-

ues of the objects are printed in boldface.

Object

Meaning

Entry range/Selection

&C

onfig

Instrument configuration

³

Ã

.

D

etector

Conductivity detector

³

Ã

.

T

hermostat

Operating temperature of conductivity cell

25, 30, 35, 40, 45°C, off

³

Ã

.

Z

eroUnit

Unit for display of auto-zero value

µS/cm

, %fs, mV

³

À

.

C

ellConst

Cell constant of conductivity cell

13.0

16.7

21.0 /cm

³

Ã.

P

rinter

External printer

³

Ã

.

I

d1

1

st

line of printout header

18 ASCII characters

³

Ã

.

Id2

2

nd

line of printout header

18 ASCII characters

³

Ã

.

P

rintHead

Output of printout header: once, always, never

once

, always, off

³

Ã

.

D

ateTime

Output of date and time in printout header

on

, off

³

À

.

S

endTo

³

Selection of printer driver

IBM

, Epson, Seiko, Citizen,

HP

³

Ã.

PrintM

easVal

³

³

Measured value printout

(function same as key)

$G, $S

³

Ã

.

P

rintCrit

³

³

Criterion for measured value printout: immediately

(single value), time (continuous output), off

immed.

, time, off

³

Ã

.

T

ime

Timed output of measured values

³

³

Ã

.

I

nterval

Time interval for output of measured values

0.4

1.0

99999 s

³

³

À

.

S

topTime

Stop time (off = infinite)

1

999 min, off

³

À

.

D

ateTime

Output of date and time of measured value

on, off