5.4 Diagnostic tests

732 IC Detector / 733 IC Separation Center

151

5.4.3

Checking working memory (RAM)

This diagnostic step executes a non-destructive test over the entire

area of the RAM content (working memory).

Prepare instrument for diagnostic test (see section 5.4.2).

If need be, press <9> key repeatedly until

diagnosis
>diag/RAM test

If no faults are found, the display shows:

>diag/RAM test
RAM test ok

diagnosis
>diag/keyboard test

5.4.4

Check keypad

With this diagnostic step all keys of the keypad are checked for func-

tionality.

Prepare instrument for diagnostic test (see section 5.4.2).

If need be, press <9> key repeatedly until

diagnosis
>diag/keyboard test

>diag/keyboard test

Press all keys in succession and check the reaction on the display.

The display shows the corresponding matrix code for the pressed key. The matrix
code can be seen in the key table (e.g. the following display should appear if the
key has been pressed).

diag/keyboard test
code: 2

The test is quit by pressing the key twice.

diagnosis
>diag/display test