6.2 Remote interfaces
732 IC Detector / 733 IC Separation Center
185
6.2.2 "733 IC Separation Center" interface
The “733 IC Separation Center” is normally attached to remote interface
14
14 "733 IC Separation Center" (see section 2.3). If the 732 IC Detector is
operated without the 733 IC Separation Center, this remote interface
can also be used for the connection of external devices. The remote
interface "733 IC Separation Center" has the following pin assignment:
732 IC Detector
Pin
Function
Inputs
+5V
21
Position "FILL" at injection valve A
L = Valve A in position "FILL"
H = Valve A not in position "FILL"
+5V
9
Position "INJECT" at injection valve A
L = Valve A in position "INJECT"
H = Valve A not in position "INJECT"
+5V
22
Position "FILL" at injection valve B or "in position" at suppressor
L = Valve B in position "FILL" or suppressor set to "in position"
H = Valve B not in position "FILL" or suppressor not set to
"in position"
+5V
10
Position "INJECT" at injection valve B
L = Valve B in position "INJECT"
H = Valve B not in position "INJECT"
+5V
+5V
23
11
2
0
2
1
Configuration of the 733 IC Separation Center
Input
Decimal
Status
11
23
code
H
H
0
undefined
H
L
1
valve A
L
H
2
valve A + suppressor
L
L
3
valve A + valve B
+5V
24
Connection 733 IC Separation Center
L = 733 IC Separation Center connected
H = 733 IC Separation Center not accessible
+5V
12
not assigned