5 Notes – Maintenance – Faults

732 IC Detector / 733 IC Separation Center

148

5.3.2

Malfunctions and their rectification

If difficulties appear with the IC system during analyses, their causes

are best investigated in the order separating column

→ pump →

eluent

→ 732/733 IC System. Several of the malfunctions which may

appear are listed in the following table with details of possible causes

and countermeasures.

Malfunction

Cause

Rectification

Baseline with high
noise level,
pulsation

Contaminated pump
values

Faulty piston seals

Quality of the pump
does not suffice for the
selected sensitivity

Clean the valves (see
section 5.1.2)

Replace the piston seals
(see section 5.1.2)

Use pulsation dampener,
use more powerful pump
or lower the sensitivity

Drift of the
baseline

Thermal equilibrium not
yet reached

Leak in system

Evaporation of organic
solvent in eluent

Condition system with
heating switched on

Check connections and
make leakproof

Ensure better closure of
eluent supply vessel

Considerable
pressure drop

Leak in system

Check connections and
make leakproof

Considerable
pressure rise

Contamination of the
filter in the 6.2821.100
Filter unit PEEK

Contamination of the
filter in the 6.2821.000
Filter unit Manufit

Contamination of the
column inlet filter

Change of column
packing by injection of
contaminated samples

Replace the 6.2821.110
Filter (see section 2.6.3)

Clean or replace
6.2821.020 Steel mesh(es)

Clean or replace
6.2821.020 Steel mesh(es)

Regenerate the column
(see section 5.1.1) or re-
place column

Note:
Samples should always
be microfiltered.

Chromatograms
with poor resolu-
tion, change in the
retention times

Deterioration in separa-
tion efficiency of the IC
column

Regenerate the column
(see section 5.1.1) or re-
place column

Extreme peak
broadening,
splitting (double
peaks)

Dead volume at the
column ends

Fill column with glass
beads (

100

µ

m) or

replace column

No feed of re-
generation or
rinsing solution
for suppressor

Contamination of the
filter in the 6.2821.100
Filter unit PEEK

Counterpressure too
high in suppressor
module

Replace the 6.2821.110
Filter (see section 2.6.3)

Clean or replace suppres-
sor module (see section
5.2.6

…5.2.8)