2 Installation

732 IC Detector / 733 IC Separation Center

22

2.6.3

Filter unit PEEK

The 6.2821.100 Filter unit PEEK (see Fig. 11) supplied with the 709

IC Pump is installed between the 709 IC Pump and the injection valve

at the 733 IC Separation Center. This filter unit serves to avoid contami-

nation of the piston seals of the 709 IC Pump by abrasive particles and

can be used in the pressure range 0

25 MPa (0

250 bar).

The two filter units PEEK supplied with the 2.733.0X30 IC Separation

Center (with suppressor) are installed between the 752 Pump Unit and

the inlet capillaries for regeneration and rinsing solution. These filter

units serve to protect the suppressor module from foreign particles and

bacterial growth.
The 6.2821.100 Filter unit PEEK consists of the housing 54

54 and the two

connectors 53

53 (with filter) and 55

55 (without filter) screwed into the hous-

ing 54

54. For the connection of capillaries 51

51 PEEK compression fittings

52

52 (6.2744.010) must be used. New connectors 53

53 with filter are avail-

able as an option with the ordering number 6.2821.110 (set of 10).

For the connection of the filter unit, please note the flow direction

arrow printed on the housing.

51

51

52

52

53

53

54

54

55

55

52

52

51

51

Fig. 11: 6.2821.100 Filter unit PEEK

51

51

Capillary

6.1831.010 PEEK capillary

54

54

Housing for filter unit

Part of 6.2824.100 Filter unit

52

52

Compression fitting (6.2744.010)

55

55

Connector without filter

Part of 6.2824.100 Filter unit

53

53

Connector with filter (6.2824.110)

Part of 6.2824.100 Filter unit