5.1 Practical notes on ion chromatography

732 IC Detector / 733 IC Separation Center

135

5.1.3

Eluents

Treatment
For the preparation of the eluents only chemicals of a purity degree of

at least "p.a." should be used. For diluting please use only high purity

water.
Fresh eluents should always be microfiltered (0.45

µ

m filter) and de-

gassed (with N

2

, He or vacuum). For high sensitive measurements, the

eluent should be continuously stirred with a magnetic stirrer, particu-

larly when the recycling procedure is employed or when alkaline eluents

are used. For alkaline eluents and eluents with low buffering capacity

one should preferably use a CO

2

absorber (e.g. the absorber supplied

with the optional 6.5324.000 Bottle rack).
The supply vessel containing the eluent must be closed as tightly as

possible to avoid excessive evaporation. This is primarily important with

eluents containing organic solvents (e.g. acetone), the evaporation of

which can lead to drifts in the long term. If work is performed in a very

sensitive range, even if one drop of condensate falls back in the eluent

this can cause a noticeable change in the background conductivity.

Influence of various parameters on anion columns

Concentration:

An increase in the concentration usually leads

to shorter retention times and quicker

separation, but also to a higher background

conductivity.

pH:

pH alterations lead to shifts in the dissociation

equilibrium and thus to changes in the

retention times.

Organic modifiers:

Addition of an organic solvent (e.g. methanol,

acetone, acetonitrile) to aqueous eluents

generally accelerates lipophilic ions.

Eluent change
When the eluent is changed, it must be ensured that no precipitates

can be formed. Solutions used in direct succession must therefore be

miscible. If the system has to be rinsed with an organic solution, several

solvents with increasing or decreasing lipophilic character may possibly

have to be used (e.g. water

acetone

chloroform).