3 Operating tutorial

732 IC Detector / 733 IC Separation Center

48

3.3

Putting into operation

After the complete IC system has been installed as described in sec-

tion 3.2, it can be put into operation. In what follows, all operating steps

up to the first calibration with the standard solution are described in se-

quence.
Please note that all displays refer to the condition in which the instru-

ment was put into operation for the first time (initial condition). If you do

not run through this tutorial until a later date, differences in regard to the

dialog language and the parameter values can appear. How you return

to the initial condition in such a case is described in section 5.4.9.
In this description of putting the instruments into operation, it is as-

sumed that all instruments are shut down and must first be switched

on. If this is not the case (e.g. if you start the operating tutorial immedi-

ately following the conditioning), you can skip steps 1

13

3.

1 Switch on external devices

Switch on recorder (if used).

Switch on «IC Metrodata for Win95» and PC

(if used).

Switch on other recording and evaluation system

(if used).

2 Switch on 709 IC Pump

Switch on 709 IC Pump with mains switch 29

29

(see section 2.6.4 of the 709 Instructions for Use).

3 Switch on remote control for 709 IC Pump

(the pump must be connected to the 732 IC Detector with a

6.2125.060 RS Cable for this)

Set sliding switch 36

36 at the rear of the 709 IC Pump to

"

RS

232

"

(see Fig. 2 of the 709 Instructions for Use).

Press key 88 on the 709 IC Pump to switch on the

external control (see section 3.6 of the 709 Instructions for

Use).
When the external control is activated, LED 77 above the

key lights up. All keys of the 709 IC Pump with the

exception of the