Metrohm 732 IC Detector User Manual
Page 182
6.1 RS232 interfaces
732 IC Detector / 733 IC Separation Center
175
Object
Meaning
Entry range/Selection
&D
iagnosis
Diagnostic tests
³
³
Ã.
E
EPROMInit
Initialize data memory
$G
³
À
.
B
lockSelect
³
³
³
³
³
³
³
³
³
³
³
³
³
Selection of blocks to be initialized:
USER
All blocks except instrument ad-
justment and instrument number
CONFIG
Block &Config
INFO
Block &Info
SETUP
Block &Setup
ASSEMBLY Block &Assembly
EVENT
Block &SelectEvent
METHOD
Blocks &Parameter and &Program
ALL
All blocks incl. instrument adjust-
ment (security code necessary)
USER
, CONFIG, INFO,
SETUP, ASSEMBLY, EVENT,
METHOD, ALL
³
Ã.
R
unMode
Parameters for sequence control
³
Ã
.
R
eport
Output of a diagnostic report on printer
on, off
³
À
.
S
top
Stop after error
on, off
³
Ã.
R
AMTest
Check working memory
$G
³
Ã.
K
eyTest
Check keypad
$G, $S
³
Ã.
D
isplayTest
Check display
$G, $S
³
Ã.
RS
Test
³
³
³
³
Check RS232 interfaces
(possible only with 3.496.8480 Test plug,
see section 5.4.6)
³
Ã
.
R
S232
Test of "RS232" interface
$G, $S
³
À
.
I
CPump
Test of "709 IC Pump" interface
$G, $S
³
Ã.
I
OTest
³
³
³
³
Check remote interfaces
(possible only with 3.496.8510 Test plug,
see section 5.4.7)
³
Ã
.
R
emote
Test of "Remote" interface
$G, $S
³
À
.
S
epCent
Test "733 IC Separation Center" interface
$G, $S
³
Ã.
I
ntTest
Internal hardware test
³
Ã
.
U
cellTest
Check voltage generator
$G, $S
³
Ã
.
R
angeTest
Check range amplifier
$G, $S
³
Ã
.
S
enseTest
Check sense amplifier
$G, $S
³
Ã
.
D
ACCoarse
Check DA converter for coarse adjustment
$G, $S
³
À
.
DACF
ine
Check DA converter for fine adjustment
$G, $S
³
Ã.
E
xtTest
External hardware test
³
Ã
.
L
owpassTest
Check low-pass function
$G, $S
³
Ã
.
O
ffsetTest
Check offset
$G, $S
³
Ã
.
P
olarTest
Check polarity
$G, $S
³
Ã
.
T
CoeffTest
Check temperature coefficient
$G, $S
³
Ã
.
F
reqTest
Check frequency
$G, $S
³
À
.
Th
ermostatTest
Check thermostat
$G, $S