4 Operation
732 IC Detector / 733 IC Separation Center
102
of the 732 IC Detector shows the message
"
stopped:
YY-MM-DD HH:MM
"
with details of the time when the
pump was stopped. The LED
"
P
min
"
on the 709 IC
Pump flashes, the shutoff pressure can be requested
under
"
P
actual
"
.
A 709 IC Pump stopped in this manner is put back into
operation as follows:
•
Press
tion.
•
Press
LED
"
P
min
"
no longer flashes.
•
Press
tion back on.
>PARAM/709 IC Pump
flow corr. 1.00
0.90
…
1.10
Correction factor of the flow rate
Here a correction factor for the flow rate can be entered
to match the displayed flow rate with the flow rate actu-
ally measured. You will find further information in sec-
tion 3.3 of the 709 Instructions for Use.
4.5.2
SCALE
FULL
The
range to avoid the circuitous route via parameter menu (see sec-
tion 4.5.1). You will find a detailed description of the parameter
"
full
scale
"
in section 4.5.1.
XXXXX
µµS/cm XXXX min
full scale: 1.00 mS/cm
0.05
µ
S/cm
…
10 mS/cm
Full-scale range (operating range)
After the
bottom line of the display shows the parameter
"f
ull
scale
"
with a flashing numeric value. In contrast to the
entry under the key, here the parameter
value is not entered using the
selected with the <ç> and <è> keys. Proceed as
follows:
•
Press the <ç> key to lower the full-scale range
stepwise to the lowest possible value.
•
Press the <è> key to increase the full-scale range
stepwise to the highest possible value.
•
Confirm the desired full-scale range with the