4 Operation

732 IC Detector / 733 IC Separation Center

102

of the 732 IC Detector shows the message

"

stopped:

YY-MM-DD HH:MM

"

with details of the time when the

pump was stopped. The LED

"

P

min

"

on the 709 IC

Pump flashes, the shutoff pressure can be requested

under

"

P

actual

"

.

A 709 IC Pump stopped in this manner is put back into

operation as follows:

Press key to switch off remote opera-

tion.

Press key to remove the block. The

LED

"

P

min

"

no longer flashes.

Press key to switch the remote opera-

tion back on.

>PARAM/709 IC Pump
flow corr. 1.00

0.90

1.10

Correction factor of the flow rate
Here a correction factor for the flow rate can be entered

to match the displayed flow rate with the flow rate actu-

ally measured. You will find further information in sec-

tion 3.3 of the 709 Instructions for Use.

4.5.2

key

SCALE

FULL

The key can be used for direct entry of the full-scale

range to avoid the circuitous route via parameter menu (see sec-

tion 4.5.1). You will find a detailed description of the parameter

"

full

scale

"

in section 4.5.1.

XXXXX

µµS/cm XXXX min

full scale: 1.00 mS/cm

0.05

µ

S/cm

10 mS/cm

Full-scale range (operating range)
After the key has been pressed, the

bottom line of the display shows the parameter

"f

ull

scale

"

with a flashing numeric value. In contrast to the

entry under the key, here the parameter

value is not entered using the