2 Installation
732 IC Detector / 733 IC Separation Center
24
2.6.5
Connection to injection valve with PEEK capillaries
For metal-free systems and for non metal-free systems in the pressure
range 0
…
25 MPa (0
…
250 bar) we recommend to use 6.1831.010
PEEK capillaries, a 6.2620.150 Pulsation dampener (see section 2.6.2)
and a 6.2821.100 Filter unit PEEK (see section 2.6.3) to connect the
709 IC Pump and the injection valve of the 733 IC Separation Center.
Proceed as follows:
For the connection of capillaries to the injection valve, use only the
6.2744.010 PEEK compression fittings. If other steel connectors are
used (e.g. 6.2620.000 and 6.2620.010), the valve connection may be
damaged!
1 Connection to 709 IC Pump
•
Cut connection capillary 22
22 (6.1831.010 PEEK capillary) to
the required length and equip with the necessary connectors.
•
Attach one end of connection capillary 22
22 to connection 23
23 of
the 709 IC Pump (see 709 Instructions for Use).
•
Attach the other end of connection capillary 22
22 to the con-
nector 53
53 (with filter) of the filter unit 64
64 (see Fig. 13).
•
Attach the PEEK capillary 65
65 cut to the required length and
equipped with connectors to the connector 55
55 of the filter unit
64
64.
2 Installation of the capillary in the IC Separation
Center
•
Unscrew the four knurled screws 37
37 of the bottom rear panel
39
39 of the 733 IC Separation Center and remove rear panel
(see Fig. 5).
•
Insert PEEK capillary 65
65 from the back into the inner com-
partment of the IC Separation Center.
•
Install rear panel 39
39 so that the capillary is positioned in
opening 40
40 „Inlet A“ or 44
44 „Inlet B“ and screw on with the four
knurled screws 37
37.
3 Connection to injection valve
•
Connect PEEK capillary 65
65 to pulsation dampener 66
66 (see
section 2.6.2). Using another PEEK capillary 65
65, connect pul-
sation dampener 67
67 to connection "5" of injection valve 68
68.
4 Mount column connection capillary
•
Connect column connection capillary 67
67 (ca. 20 cm of
6.1831.010 PEEK capillary) to connection "4" of injection valve
68
68.