2 Installation

732 IC Detector / 733 IC Separation Center

24

2.6.5

Connection to injection valve with PEEK capillaries

For metal-free systems and for non metal-free systems in the pressure

range 0

25 MPa (0

250 bar) we recommend to use 6.1831.010

PEEK capillaries, a 6.2620.150 Pulsation dampener (see section 2.6.2)

and a 6.2821.100 Filter unit PEEK (see section 2.6.3) to connect the

709 IC Pump and the injection valve of the 733 IC Separation Center.

Proceed as follows:

For the connection of capillaries to the injection valve, use only the

6.2744.010 PEEK compression fittings. If other steel connectors are

used (e.g. 6.2620.000 and 6.2620.010), the valve connection may be

damaged!

1 Connection to 709 IC Pump

Cut connection capillary 22

22 (6.1831.010 PEEK capillary) to

the required length and equip with the necessary connectors.

Attach one end of connection capillary 22

22 to connection 23

23 of

the 709 IC Pump (see 709 Instructions for Use).

Attach the other end of connection capillary 22

22 to the con-

nector 53

53 (with filter) of the filter unit 64

64 (see Fig. 13).

Attach the PEEK capillary 65

65 cut to the required length and

equipped with connectors to the connector 55

55 of the filter unit

64

64.

2 Installation of the capillary in the IC Separation

Center

Unscrew the four knurled screws 37

37 of the bottom rear panel

39

39 of the 733 IC Separation Center and remove rear panel

(see Fig. 5).

Insert PEEK capillary 65

65 from the back into the inner com-

partment of the IC Separation Center.

Install rear panel 39

39 so that the capillary is positioned in

opening 40

40 „Inlet A“ or 44

44 „Inlet B“ and screw on with the four

knurled screws 37

37.

3 Connection to injection valve

Connect PEEK capillary 65

65 to pulsation dampener 66

66 (see

section 2.6.2). Using another PEEK capillary 65

65, connect pul-

sation dampener 67

67 to connection "5" of injection valve 68

68.

4 Mount column connection capillary

Connect column connection capillary 67

67 (ca. 20 cm of

6.1831.010 PEEK capillary) to connection "4" of injection valve

68

68.