4.8 Data output
732 IC Detector / 733 IC Separation Center
121
Automatic measured value output
•
For the automatic measured value output of measured values at se-
lectable time intervals,
"
>CONFIG/print meas.value/print crit.
=
time
"
must be selected (see section 4.4.2). When the
is pressed, the measured value output defined by the parameters
"
time interval
"
and
"
stop time
"
is started via the RS232 interval.
The measured value output can always be aborted with
•
Example:
Automatic measured value output with single print
header output (with date and time), run time,
measured value
>CONFIG/printer
id.1 column PRP-X100
id.2 conditioning
print header: once
date&time: on
send to: Seiko
>CONFIG/print meas.value
print crit.: time
time interval 600.0 s
stop time 60 min
date&time: off
732 IC Detector 01104 732.0012
Print header
id.1 column PRP-X100
Identification 1
id.2 conditioning
Identification 2
date 1995-10-09 time 14:36:43
Date and time
0.0 109.6 uS/cm
Run time, 1
st
meas. value
10.0 125.1 uS/cm
Run time, 2
nd
meas. value
20.0 133.9 uS/cm
Run time, 3
rd
meas. value
30.0 138.2 uS/cm
Run time, 4
th
meas. value
40.0 140.6 uS/cm
Run time, 5
th
meas. value
50.0 141.0 uS/cm
Run time, 6
th
meas. value
60.0 141.2 uS/cm
Run time, 7
th
meas. value