121 automatic measured value output – Metrohm 732 IC Detector User Manual

Page 128

background image

4.8 Data output

732 IC Detector / 733 IC Separation Center

121

Automatic measured value output

For the automatic measured value output of measured values at se-

lectable time intervals,

"

>CONFIG/print meas.value/print crit.

=

time

"

must be selected (see section 4.4.2). When the key

is pressed, the measured value output defined by the parameters

"

time interval

"

and

"

stop time

"

is started via the RS232 interval.

The measured value output can always be aborted with .

Example:

Automatic measured value output with single print

header output (with date and time), run time,

measured value

>CONFIG/printer
id.1 column PRP-X100
id.2 conditioning
print header: once
date&time: on
send to: Seiko
>CONFIG/print meas.value
print crit.: time
time interval 600.0 s
stop time 60 min
date&time: off

732 IC Detector 01104 732.0012

Print header

id.1 column PRP-X100

Identification 1

id.2 conditioning

Identification 2

date 1995-10-09 time 14:36:43

Date and time

0.0 109.6 uS/cm

Run time, 1

st

meas. value

10.0 125.1 uS/cm

Run time, 2

nd

meas. value

20.0 133.9 uS/cm

Run time, 3

rd

meas. value

30.0 138.2 uS/cm

Run time, 4

th

meas. value

40.0 140.6 uS/cm

Run time, 5

th

meas. value

50.0 141.0 uS/cm

Run time, 6

th

meas. value

60.0 141.2 uS/cm

Run time, 7

th

meas. value