1 Introduction

732 IC Detector / 733 IC Separation Center

10

1.4

Safety notes

1.4.1

Electrical safety

While electrical safety in the handling of the 732 IC Detector and 733

Separation Center is assured in the context of the specifications IEC

1010-1 (protection class 1, degree of protection IP40), the following

points should be noted:

Mains connection

Setting of the mains voltage, checking the mains fuse and the

mains connection must be effected in accordance with the instruc-

tions in section 2.4.

Opening the 732 IC Detector

If the 732 IC Detector is connected to the power supply, the instru-

ment must not be opened nor must parts be removed from it, other-

wise there is a danger of coming into contact with components which

are live. Hence, always disconnect the instrument from all voltage

sources before you open it and ensure that the mains cable is

disconnected from mains connection 18

18 !

Opening the 733 IC Separation Center

Disconnect connecting cable to the 732 IC Detector from con-

nector 14

14 before you remove the middle housing panel with con-

nectors.

Protection against static charges

Electronic components are sensitive to static charging and can be

destroyed by discharges. Before you touch any of the components

inside the 732 IC Detector or 733 IC Separation Center, you should

earth yourself and any tools you are using by touching an earthed

object (e.g. housing of the instrument or a radiator) to eliminate any

static charges which exist.

1.4.2

General precautionary rules

Handling of solvents

Check all lines of the IC system periodically for possible leaks. Follow

the relevant instructions regarding the handling of flammable and/or

toxic solvents and their disposal.