beautypg.com

Metrohm 732 IC Detector User Manual

Page 44

background image

2.8 Separating columns and suppressor module

732 IC Detector / 733 IC Separation Center

37

91

91

Suppressor outlet capillary for

H

2

SO

4

94

94

Suppressor outlet capillary

for eluent

92

92

Suppressor outlet capillary for

H

2

O

95

95

Suppressor module

93

93

Suppressor inlet capillary for

H

2

O

96

96

Coupling (6.2620.060; steel) or

Coupling (6.2744.040; PEEK)

2.8.6

One-channel system with suppressor module

In the case of the one-channel system with suppressor module, first the

IC separating column is installed in the 2.733.0X30 IC Separation Cen-

ter (see Fig. 17) and then the suppressor module is connected to the

752 Pump Unit needed for operation (see Fig. 18). Proceed as follows:

1 Connect column to injector

Remove end caps from column 76

76.

without precolumn:

Screw inlet end of separating column 76

76 (note flow direction)

to column connection capillary 67

67 mounted on the injector.

with precolumn in twin cartridge holder:

Screw inlet end of separating column 76

76 (note flow direction)

to outlet capillary 72

72 (see Fig. 15).

with precolumn in cartridge head:

Install separating column 76

76 (note flow direction) in the car-

tridge head as described in section 2.7.3 (see Fig. 15).

2 Rinse column

Place a beaker beneath the column outlet.

Switch on 709 IC Pump and rinse column with eluent for ca.

10 min.

Switch off IC pump.

3 Connect column to suppressor module

Cut inlet capillary 89

89 marked with "Eluent" of suppressor

module 95

95 to the desired length using a sharp cutting tool

(e.g. razor blade).

Screw inlet capillary 89

89 to outlet end of separating column 76

76

using a 6.2744.010 Compression fitting.

4 Fix column

Insert one or two column holders 84

84 (6.2027.030, 6.2027.040

or 6.2027.050) in mounting rails 83

83 and fasten separating

column in the column holder.

5 Connect suppressor module to detector block

Cut outlet capillary 94

94 marked with "Detector" of suppressor

module 95

95 to the desired length using a sharp cutting tool

(e.g. razor blade).