Metrohm 732 IC Detector User Manual
Page 67
![background image](/manuals/586520/67/background.png)
3 Operating tuturial
732 IC Detector / 733 IC Separation Center
60
OVERLOAD
As the selected full-scale range is clearly less than the
measured absolute conductivity, red overload display 88 on
the 732 IC Detector now lights up.
3 Trigger auto-zero
ZERO
⇒
Press the
tronic background compensation. During zeroing, the
green LED in the key starts to flash.
+0.000
µµS/cm 56.4 min
full scale: 4.00
µµS/cm
As soon as zeroing is at an end, the green LED in the
shows the auto-zero value in place of the absolute con-
ductivity. In contrast to the absolute conductivity, the auto-
zero value is always outputted with the sign in front of the
number.
With the auto-zero function the current conductivity value
becomes the new zero point of the selected full-scale
range. Each time the
uring signal is always reset to zero. The auto-zero function
can be switched off with the
4 Set injection valve A to "FILL"
FILL
⇒
Press the
Separation Center. This switches the injection valve to the
"FILL" position. This position is indicated by lighting up of
the green LED in the
5 Fill sample loop
⇒
Immerse the aspirating tubing 88
88 attached to connection
22
22 in the standard solution.
⇒
Using the syringe fixed to connection 21
21 siphon in ca. 1 mL
standard solution.
6 Set injection valve A to "INJECT"
INJECT
⇒
Press the
Separation Center and the chromatogram is started. This
switches the injection valve to the "INJECT" position. The
"INJECT" position is indicated by lighting up of the green
LED in the
⇒
At the same time start the recorder (if used).
An integrator attached to the 733 IC Separation Center or a
PC with evaluation software (e.g. «IC Metrodata for Win95»)
is started automatically when injection valve A is switched
to the "INJECT" position.