3 Operating tutorial
732 IC Detector / 733 IC Separation Center
56
>PARAM/analog output
damping: off
The electronic damping of the analog output signal must
normally not be switched on.
ENTER
⇒
Confirm the switched-off status of the damping with the
PARAM
>PARAM/plot
The title of the submenu
"
plot
"
appears. This submenu
contains various inquiries regarding the graphics printout
on an external printer. As we shall not deal with the settings
for such a printer in this brief tutorial, we can switch directly
to the next inquiry.
9 Enter parameters for 709 IC Pump
If you do not operate the 709 IC Pump by remote control
via the 732 IC Detector, the parameters below must be
entered directly on the pump (see 709 Instructions for
Use).
PARAM
⇒
Press the key.
PARAM
>PARAM/709 IC Pump
The submenu
"
709 IC Pump
"
contains various inquires
concerning the 709 IC Pump.
ENTER
⇒
Press the
>PARAM/709 IC Pump
flow 0.5 mL/min
The first inquiry concerns the flow of the 709 IC Pump,
which can be set to a value between 0.01 and 5.00 mL/min
using the numeric keys.
>PARAM/709 IC Pump
flow 2.0 mL/min
For the PRP-X100 IC Anion Column used in the example,
the recommended flow rate is 2 mL/min. Enter this value
using the numeric keys.
ENTER
⇒
Confirm the inputted value using the
>PARAM/709 IC Pump
Pmax 10.0 MPa
The parameter
"
Pmax
"
denotes the maximum shutoff
pressure for the 709 IC Pump and is used to protect the
separating column against excessive pressure. If the pres-
sure exceeds this value, the pump is automatically shut off.
Use the numeric keys to enter the desired value here. This
limit value should be ca. 3 MPa above the normal operating
pressure of the separating column used or correspond to
the maximum admissible operating pressure of the column
(1 MPa = 10 bar).