4 Operation

732 IC Detector / 733 IC Separation Center

104

4.6.2

keys

The two keys at the 733 IC Separation Center have the

following function, depending on the instrument version:

733.0010 (1 valve)

733.0X20 (2 valves)

733.0X30 (valve+ suppr.)

INJECT

A

Switching of injection

valve A to "INJECT"

position

Switching of injection

valve A to "INJECT"

position

Switching of injection

valve A to "INJECT"

position

INJECT

B

No function

Switching of injection

valve B to "INJECT"

position

No function

If the injection valve is in the

"INJECT" position, the green LED

in the key lights up. In this posi-

tion, the injection valve is con-

nected as shown in the diagram

opposite:

The eluent flows via the sample

loop to the separating column.

This causes injection of the

sample previously added to

the sample loop.

Sample inlet tubing and sy-

ringe are directly connected.

The "INJECT" function can be triggered at any time with the

key – also in the edit mode or when a program is running. If this is not

required, the key can be locked. For this, the parameter

"

>CONFIG/733

IC

Sep.Cent./control

"

must be set to

"

732 only

"

under the

key (see section 4.4.2). Switching of the injection valve is

then possible only using the 732 IC Detector via a programmed

"INJECT" command or with

"

>CONFIG/733

IC

Sep.Cent./trigger = in-

ject

"

. Switching of injection valve A to the "INJECT" position can also

be used for the automatic start of a program of the "inject" type.
With the 733.0X30 Separation Center, switching of the suppressor

module to the next position can be triggered automatically by the

switching of injection valve A to the "INJECT" position. For this, the pa-

rameter

"

>CONFIG/733/suppressor/auto step

"

must be set to

"

inject

"

under the key (see section 4.4.2).

1

2

4

5

3

6

Syringe

Column

Eluent

Sample

Loop