beautypg.com

6 interfaces – Metrohm 732 IC Detector User Manual

Page 191

background image

6 Interfaces

732 IC Detector / 733 IC Separation Center

184

732 IC Detector

Pin

Function

Outputs

5

Ready
L = ready (program inactive or not started)
H = not ready (program running)

V

CEO

= 40 V

IC

= 20 mA

18

Remote 1
Programmable remote output line 1
(command "remote" in program or event)

V

CEO

= 40 V

IC

= 20 mA

4

Remote 2
Programmable remote output line 2
(command "remote" in program or event)

V

CEO

= 40 V

IC

= 20 mA

17

Advance pulse (EOD)
The EOD signal is outputted in the following cases:

without program: on every "INJECT A"

with program: each time a program is ended and
on every return to the loop start with loop programs

H

L

t

P

> 100 ms

3

Remote 3
Programmable remote output line 3
(command "remote" in program or event)

V

CEO

= 40 V

IC

= 20 mA

16

Error
L = instrument error (as long as it exists)
H = instrument ok

V

CEO

= 40 V

IC

= 20 mA

1

Remote 4
Programmable remote output line 4
(command "remote" in program or event)

V

CEO

= 40 V

IC

= 20 mA

2

Remote 5
Programmable remote output line 5
(command "remote" in program or event)

V

CEO

= 40 V

IC

= 20 mA

Voltages

+5V

0V

15

14

25

5 V: inactive High (H)
0 V: active

Low (L)

I

40 mA

R

i

12

Contact arrangement at socket "Remote"
(female)

13

1

25

14

Connecting cables to external devices must have
an appropriate 25-pin connector (male). Metrohm
offers connector sockets (K.210.9060) and the
associated housing (K.210.0002) as an option for
the preparation of such cables.

No liability whatsoever will be accepted for damage arising from the improper connection of devices.