2.5 Capillary connections

732 IC Detector / 733 IC Separation Center

19

2.5

Capillary connections

2.5.1

Capillaries

Some of the connections under high pressure between the feed pump

and the detector block must be set up by the user. For metal-free sys-

tems, the 6.1831.010 PEEK capillary (i.d. = 0.25 mm, e.d. =

1

/

16

"

,

length = 3 m) must be used.
For non metal-free systems, the 6.1831.010 PEEK capillary can also

be used in the pressure range of 0

25 MPa (0

250 bar); on the other

hand, in the pressure range of 25

50 MPa (250

500 bar), which is

permissible only together with the non metal-free 2.709.0010 version of

the 709 IC Pump, the 6.2620.020 Steel capillary (i.d. = 0.25 mm,

e.d. =

1

/

16

"

, length = 3 m) must be used.

With PEEK capillaries the connection is made preferably with the

6.2744.010 PEEK connectors, with steel capillaries, the connection is

made preferably with 6.2620.000 and 6.2620.010 steel connectors (see

section 2.5.2 and section 2.5.3).

Capillaries fitted with new connectors must have a perfectly flat cut

surface. For PEEK capillaries it is best to use the 6.2621.080 Capil-

lary tubing cutter, for steel capillaries the 6.2621.040 Capillary

tubing cutter (both available as an option).

2.5.2

Steel connectors

For the connection of steel capillaries, the steel connectors 6.2620.010

Ferrule and 6.2620.000 Pressure screw available as an option can

be used. Proceed as follows:

1 Mount connectors

Slide a pressure screw 50

50 (6.2620.000) and a ferrule 49

49

(6.2620.010) over the end of the capillary 51

51 to be fastened as

shown in Fig. 9.

2 Insert capillary in connection

Push capillary end into the corresponding connection as far as it

will go (to avoid dead volume).

3 Tighten compression fitting

Tighten pressure screw 50

50 with the open-end spanner

1

/

4

"

(6.2621.050) supplied.

For the connection of capillaries to the injection valves, use only the

special steel connections contained in a plastic bag affixed to the

valve (or as an alternative the 6.2744.010 PEEK compression fittings).

If other steel connectors are used (e.g. 6.2620.000 and 6.2620.010),

the valve connection may be damaged!