beautypg.com

Metrohm 732 IC Detector User Manual

Page 97

background image

4 Operation

732 IC Detector / 733 IC Separation Center

90

>CONFIG/printer
id.2

18 ASCII characters

Identification 2 for 2

nd

line of print header

Freely selectable ASCII character string for the second

line of the print header (text entry, see section 4.2.5).

>CONFIG/printer
print header: once

once, always, off

Printout of the print header
The print header comprises title line (with device name,

serial number and program number), identification 1

and 2 as well as date and time if selected and is printed

out before a measured value, report or curve plot.

once

Printout of the print header only once

after the instrument has been switched

on before each measured value, report

or curve plot.

always

Printout of the print header before every

measured value, report or curve plot.

off

No printout of the print header.

>CONFIG/printer
date&time: on

on, off

Printout of date and time in the print
header

on

Date and time are printed out in the print

header.

off

Date and time are not printed out in the

print header.

>CONFIG/printer
send to: IBM

IBM, Epson, Seiko,

Citizen, HP

Selection of the character set/printer type

IBM

Character set 437 for IBM PC or printer

type IBM Proprinter

Epson

EPSON printers

Seiko

Seiko printer DPU-411/414

Citizen

Citizen printer IDP562 RS

HP

HP printers (Deskjet..., Laserjet... etc.)

You will find further details on the printer connection in

section 2.9.5.

>CONFIG
>CONFIG/print meas.value

Settings for measured value printout
Output of measured values via RS232 interface.

>CONFIG/print meas.value
print crit.: immed.

immed., time, off

Criterion for measured value printout

immed.

Measured value printout each time the

is pressed.

time

Automatic measured value printout at

selectable time intervals.

off

No measured value printout.