4.8 Data output
732 IC Detector / 733 IC Separation Center
119
4.8
Data output
4.8.1
3
The
to start the automatic measured value output on an external printer or
PC connected to the RS232 interface. The output is triggered directly at
a keystroke, no inquiries appear.
Preparations
•
For the output on an external printer, printer type and RS232 set-
tings must be correctly selected under
tion 4.4.2).
•
For the output on a PC the character set under
"send
to: IBM"
) and the RS232 settings must be correctly selected see
section 4.4.2). For receipt of the data, a suitable program must be
started on the PC (e.g. Metrodata VESUV 2.1 or terminal program in
Windows accessories).
Elements of the measured value output
Depending on the parameter settings under
"
>CONFIG/print meas.
value
"
, the measured value output can comprise the following ele-
ments:
•
Print header
The print header comprises a title line (device name, serial number,
program number), identifications 1 and 2 (if used) as well as the
date and time if desired (see section 4.4.2) and is outputted once,
always or never before a measured value output, depending on the
setting of the parameter
"
>CONFIG/printer/print header
"
.
•
Run number
The run number (parameter
"
>CONFIG/auxiliaries/run number
"
,
see section 4.4.2) is incremented by +1 and outputted on every
"INJECT A" (without program) or on every new start or
"
return
"
(with
program).
•
Run time
The current time after the program start, after the last "INJECT A" or
after the instrument has been switched on is outputted before the
measured value.
•
Measured value
The measured value shown in the display is outputted:
Measured value without sign
Absolute conductivity
in
µ
S/cm or mS/cm
Measured value with sign
Auto-zero value in
µ
S/cm,
%fs or mV
"--------"
Measured value is >10%
outside the measuring range