4 Bedienung
732 IC Detector / 733 IC Separation Center
82
4.3.2
Status messages
SELECT
Pressing the
messages in turn. These are shown in the bottom line of the display.
Depending on the instrument status, the following messages can be
displayed:
+0.004
µµS/cm
13.5 min
method XXXXXXXX
Status messages for a method
method XXXXXXXX
An unchanged method is loaded in the
working memory (see section 4.7.4).
method XXXXXXXX modif.
The method loaded in the working memory
has been changed (see section 4.7.4).
method DEFAULT
The default method
"
DEFAULT
"
is loaded in
the working memory (see section 4.7.4).
+0.004
µµS/cm
13.5 min
1995-10-12 11:24:31
Display of date and time
The current date and the current time are dis-
played. Date and time can be changed under
151.2
µµS/cm
13.5 min
full scale 1.00 mS/cm
Display of the full-scale range
The full-scale range currently set is displayed
(operating range, see section 4.5.1) and can be
changed with the
section 4.5.2)
+0.004
µµS/cm
13.5 min
abs.cond. 151.2
µµS/cm
Display of the absolute conductivity
The absolute conductivity currently measured is
displayed. This is primarily useful when the auto-
zero value is displayed as the measured value.
+0.004
µµS/cm
13.5 min
pump ready
Status messages for 709 IC Pump
These status messages appear only with
"
SETUP/peripherals/operation with 709 = on
"
.
pump ready
The 709 IC Pump is switched on and ready
for start with the
section 4.6.6).
pump running XX.X MPa
The 709 IC Pump is running, the current
pressure P
actual
is shown on the right. The
pump can be stopped using the R/S> key (see section 4.6.6).