4.3 Displays in the standby mode
732 IC Detector / 733 IC Separation Center
83
stopped: YY-MM-DD HH:MM
The 709 IC Pump was stopped at the time
indicated owing to violation of the shutoff
limit
"
Pmin
"
or
"
Pmax
"
. To restart a pump
stopped in this manner, the pump keys
succession and then
again (see also section 4.5.1).
pump not responding
This message appears when
•
the 709 IC Pump is not connected to the
732 IC Detector,
•
the 709 IC Pump is not switched on,
•
the remote control of the 709 IC Pump is
not switched on.
+0.004
µµS/cm
13.5 min
prog.type cycle no. XXX
Status messages for program type
These status messages appear only with
"
PROGRAM/parameters/status = active
"
.
prog.type cycle no. XXX
Display of the program type
"
cycle
"
and the
cycle number (see section 4.7.1).
prog.type inject
Display of the program type
"
inject
"
(see
section 4.7.1).
prog.type remote
Display of the program type "
remote
" (see
section 4.7.1).
+0.004
µµS/cm
13.5 min
next step ready
Status messages for program step
These status messages appear only with
"
PROGRAM/parameters/status = active
"
.
next step ready
The program is ready for a new start (see
section 4.7.1).
next step XXX.X min
Display of the time when the next program
step will be executed.
next step ---
The program has been set to
"
inactive
"
, a
new start is not possible.
next step
ì
The program continues to run indefinitely (no
"
end
"