beautypg.com

Metrohm 732 IC Detector User Manual

Page 90

background image

4.3 Displays in the standby mode

732 IC Detector / 733 IC Separation Center

83

stopped: YY-MM-DD HH:MM

The 709 IC Pump was stopped at the time

indicated owing to violation of the shutoff

limit

"

Pmin

"

or

"

Pmax

"

. To restart a pump

stopped in this manner, the pump keys

and must be pressed in

succession and then pressed

again (see also section 4.5.1).

pump not responding

This message appears when

the 709 IC Pump is not connected to the

732 IC Detector,

the 709 IC Pump is not switched on,

the remote control of the 709 IC Pump is

not switched on.

+0.004

µµS/cm

13.5 min

prog.type cycle no. XXX

Status messages for program type

These status messages appear only with
"

PROGRAM/parameters/status = active

"

.

prog.type cycle no. XXX

Display of the program type

"

cycle

"

and the

cycle number (see section 4.7.1).

prog.type inject

Display of the program type

"

inject

"

(see

section 4.7.1).

prog.type remote

Display of the program type "

remote

" (see

section 4.7.1).

+0.004

µµS/cm

13.5 min

next step ready

Status messages for program step

These status messages appear only with
"

PROGRAM/parameters/status = active

"

.

next step ready

The program is ready for a new start (see

section 4.7.1).

next step XXX.X min

Display of the time when the next program

step will be executed.

next step ---

The program has been set to

"

inactive

"

, a

new start is not possible.

next step

ì

The program continues to run indefinitely (no

"

end

"

command available).