3 Operating tutorial

732 IC Detector / 733 IC Separation Center

46

6.2821.100 Filter unit PEEK

Filter between 709 IC Pump and injection valve

to avoid contamination.

6.1005.000 PRP-X100 IC Anion Column

Eluent

2 mmol/L phthalic acid / 8% acetone / pH 5.0 in dist. H

2

O

flow: 2 mL/min

Standard

Standard solution with 5 ppm Cl

and

10 ppm each of NO

3

and SO

4

2–

(in dist. H

2

O)

2.714.0310 IC Metrodata for Win95

The «IC Metrodata for Win95» PC integration system, which com-

prises a PC board and the associated PC integration software, is

used to record and evaluate chromatograms on a PC. The in-

stallation, operation and method development are not described

in this brief tutorial. You will find all relevant information in the In-

structions for Use of IC Metrodata for Win95.
Instead of «IC Metrodata for Win95», you can also use other data

recording systems, integrators, recorders or printers, but their use

is also not described in this tutorial.

3.2

Preparations

Before you start this brief tutorial, the entire IC system must be correctly

installed as described in section 2. In what follows, the most important

points for the installation are described once again (for details, see the

sections mentioned).

1 Install 732 IC Detector and 733 IC Separation

Center

Setting up instruments

section 2.2

Installing and connecting detector block

section 2.3.1/2

Mounting syringe and aspirating tubing

section 2.3.3

Mounting drain tube

section 2.3.4

Power supply connection

section 2.4

2 Prepare eluent

Preparing eluent:

2 mmol/L phthalic acid / 8% acetone / pH 5.0 in dist. H

2

O

(pH value of solution adjusted with NaOH)

Microfiltering and degassing eluent

section 5.1.3

Stirring eluent in supply vessel

section 5.1.3