6 Interfaces
732 IC Detector / 733 IC Separation Center
164
6.1.4
Status messages
To ensure appropriate control of an external control unit is possible, it
must also be possible to request status conditions which provide in-
formation on the status of the 732 IC Detector. The output of a status
message is triggered by the trigger
'$D'.
Status messages comprise
the global status
'$R'
and the detailed status messages listed below,
which can also appear in combination, e.g.
'$R.Cond.ProgWait'
. If a
fault appears, the error message is also appended to the status mes-
sage, e.g.
'$R.Cond.ZeroOK;E42'
.
$R
Ready: The 732 IC Detector is ready to measure.
.Cond
Basic status of the conductivity measurement
.Zero
Auto-zero function active, value not yet valid
.ZeroOK
Auto-zero function active, value valid
.Overflow
The measured value is 50
…
80% outside the full-scale range
.Background
The measured value is more than 80% outside the full-scale
range
.ProgWait
Program active (can be started)
.ProgGo
Program running
.ProgEnd
Program ended
.ProgTest
Program test mode active
.Input
Change in the signal of an input line
.Output
Change in the signal of an output line
.Diagnosis
Instrument in diagnostic status